ForTest - Leak test and Flow test equipment

Tel. +39 0535 731530email info@fortest.com
ForTest Home > ProductsLeak Test > Compliance > Compliance/Interception leak tester standard level M8980

Compliance/Interception leak tester standard level M8980


  • Born for testing under hermetic bell
  • Almost immediate test result
  • Test Pressure up to 10 Bar

application fields

M8980 is the new Leak test Compliance/interception of ForTest® M series.

This equipment  detects leaks in  the external working environment of the piece to be tested, and it is, therefore, applicable to any type of component provided that there is a possibility to wrap itself in a hermetic bell or  the possibility to intercept the leakage in two different chambers (generally valves).

The test runs entering a pressure in the piece from the inside part or from one of the two chambers,  then, pressure changes will be detected in the environment between the hermetic bell and the piece or in the other chamber. The revolutionary Flash technology with ultra fast elaboration times allows to have a graphic display with Cartesian coordinate plane and with auto scale of the leak test.
M8980 has a circuit of self-diagnosis for testing bell air-tightness.

This equipment is available with test pressure from -1 to 10 Bar.

M8980 has got a 24 Vdc Aux connector for expansion signals, a USB port, a RS232 opto isolated port and a opto isolated port for industrial multi test stations networks. We can supply also external accessories as: printers, Staubli fittings for the connection to the front panel of M0710 (calibrated leaks), kits network sorting for the realization of complex PC networks, mechanical precise regulators  etc …

If requested, it is possible to install a USB flash drive to record the tests, radio remote control to manage the tests, Bluetooth, Ethernet and Internet direct connections.

Thanks to its renewed hardware, M8980 shows a measurement accuracy of 0,01 Pa which allows to make tests at the highest quickness and reliability possible in air leak test field. The equipment also has got: a very simple user interface, multi language menu, bar graph indicator which shows test pressure and the detected decay, RGB led  to show good or bad result.   

This testing method is characterized by a very high sensitivity (for legislation 10/100 times than conventional systems, and works very well at high pressure) and by  testing speed, as the same coincides with the phase of pressurization of particular to be tested.
 

 

 


Tecnical Data

- Black anodized aluminium case
- “No break-through” plexiglass front panel
- Display of test pressure and of pressure decay in:  mBar / mBar/s / Pa / hPA / cc/min / cc/h
- LCD display: 240x64 pixels and blue back-illumination
- Graphic visualization of the leak test on Cartesian coordinate plane
- Mechanical pressure regulator
- Electronic pressure regulator on demand
- Measurement Accuracy better than 0.01 Pa
- Pieces Counter and Real time Clock Memory
- Possibility to set up till 100 test programs
- Calibration according to ISO rules (Menu accessible to the user and concealed by a control algorithm)
- Connector with 24 Vdc expansion signals
- RS232 opto isolated port
- Opto isolated Port for industrial multi points networks
- USB port for PC connection
- USB flash drive to record the tests, radio remote control to manage the tests, Bluetooth, Ethernet and Internet direct connections (on request)
- Working temperature min. 5°C / max. 40 °C
- Weight for the basic version: 8 Kg.
- Dimensions: 270X160X300 mm.

- Pressure full scale: (mbar) -1000; 200; 500; 1000; 2000; 3000; 6000; 10000
- Power supply: 110 Vac 50/60Hz,230 Vac 50/60Hz
- Different pneumatics available for tubes (mm): 4x2,7,6x4,8x6
- Test Gas: Hidrogen,Refrigerant,Compressed air,Nitrogen,Helium 

Optionals 

 
Interior exhaust valve
Frontal precision pressure regulator
Pneumatic module of remote measurements
Electronic pressure Regulator
Auto zero primary pressure circuit
Pre-Filling
Waiting Primary pressure reading - (INCLUDED IN STANDARD MODEL) -
Secondary Output of Waste (third result)
Leak Analog Output
I/O 24 Vdc expansion signal card
Bi-manual logic Start Input
Ethernet plug/TCP-IP
Active USB port
Bluetooth connection
Radio remote control selection Kit
High-Resolution measure (1M pti AD) - (DI SERIE) -
Capacitance meter for piece volume measure - (DI SERIE) -
Automatic Double branch pre-regulated pressures
Indirect Measure - (DI SERIE) -
 

 

Accessories 

 
External exhaust solenoid
External start button
External start/abort pushbutton
External pushbutton 4 programmes selection
Start pedal
Barcod reader
Thermal printer
Adhesive label for printer 4500pcs
Aux signals extension cable
Venturi Vacuum generator
Air filter
Staubli male connector
Precision micrometer nozzle
Micrometer nozzle
M series manager software
TTY
Serial protocol
 

M8980
Click to enlarge